5½ digit multifunction calibrator DC calibration

PT scheme / program name: Electrical calibration
PT item type: AC/DC Voltage, AC/DC Current, Resistance
Sample / item: 5½ digit multifunction calibrator DC
Model: Meatest M143
Reference Laboratory: AG Metrology S.r.l.
Measurement points and Reference Laboratory CMC:

Type of Test CMC Measurement Points
IN DCV RANGE
10 mV 0.0011 % rdg + 0.49 µV (-10, -1, 1, 10) mV
100 mV 0.0011 % rdg + 0.49 µV (-100, -10, 10, 100) mV
1 V 0.00083 % rdg + 1.8 µV (-1.0, -0.1, 0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.7, 0.8, 0.9, 1.0) V
10 V 0.00072 % rdg + 6.3 µV (-10, -1, 1, 10) V
100 V 0.00083 % rdg + 0.12 mV (-100, -10, 10, 100) V
1000 V 0.0011 % rdg + 2.1 mV (-1000, -100, 100, 1000) V
IN DCI RANGE
200 µA 0.0051 % rdg + 1.3 nA up to 100 µA

0.0051 % rdg + 7.9 nA above 100 µA

(-200, -20, 20, 200) µA
2 mA 0.0051 % rdg + 7.9 nA up to 1 mA

0.0051 % rdg + 79 nA above 1 mA

(-2.0, -0.2, 0.2, 2.0) mA
20 mA 0.0051 % rdg + 79 nA up to 10 mA

0.011 % rdg + 1.3 µA above 10 mA

(-20, -2. 2, 20) mA
200 mA 0.011 % rdg + 1.3 µA up to 100 mA

0.016 % rdg + 24 µA above 100 mA

(-200, -20, 20, 200) mA
2 A 0.016 % rdg + 24 µA up to 1 A

0.0041 % rdg + 0.11 mA above 1 A

(-2.0, -0.2, 0.2, 2.0) A
20 A 0.0041 % rdg + 0.11 mA (-20, -2, 2, 20) A
IN RESISTANCE RANGE
10 Ω 0.011 % rdg + 0.11 mΩ 10 Ω
100 Ω 0.0029 % rdg + 0.42 mΩ 100 Ω
1.0 kΩ 0.0026 % rdg + 4.2 mΩ 1.0 kΩ
10 kΩ 0.0026 % rdg + 42 mΩ 10 kΩ
100 kΩ 0.0026 % rdg + 0.42 Ω 100 kΩ
1 MΩ 0.0036 % rdg + 5.4 Ω 1.0 MΩ
10 MΩ 0.021 % rdg + 0.23 kΩ 10 MΩ
100 MΩ 0.073 % rdg + 15 kΩ 100 MΩ