Loading Events

Fault loop impedance with high test current meter

PT scheme / program name: Electrical calibration
PT item type: Electrical safety
Sample / item: insulation resistance tester
Model / type: AEMC 5050
Reference laboratory: ISO IEC 17025 accreditated laboratory or NMI

Measurement points and Reference Laboratory CMC:

Measurement Points CMC
(0.1, 0.5, 1.0, 5.0, 10, 100, 190) MΩ 50 V 0.3 % rdg
( 0.5, 1.0, 5.0, 10, 100, 190) MΩ 500 V 0.3 % rdg
(1, 2, 3, 5, 10, 15, 20, 25, 30) MΩ 1000 V 1.0 % rdg
(1, 10, 100, 1000) GΩ 5000 V 2.0 % rdg

Note: Indications regarding organizational methods, presentation and sending of results will be reported in the PT participation protocol

Scheduled start date for PT: every year in September

Payment: bank transfer

Go to Top