Temperature calibrator calibration
EPTIS ID: 703492
PT scheme / program name: Electrical calibration
PT item type: AC/DC Voltage, AC/DC Current, Resistance
Sample / item: temperature calibrator
Model / type: Beamex MC2-TE
Reference Laboratory: AG Metrology S.r.l.
Measurement point and Reference Laboratory CMC:
Type of Test | CMC | Measurement Points |
---|---|---|
Temperature Calibration, Indication and Control Equipment used with Thermocouple Type S | 0.11 – 0.0022 % rdg °C with 0° C RJ 0.17 °C with internal RJ |
(399, 700, 1000, 1300, 1700) °C |
Temperature Simulation use with Thermocouple Type S | 0.032 °C + 0.0012 % rdg with 0° C RJ 0.14 °C with internal RJ |
(399, 700, 1000, 1300, 1700) °C |
Temperature Calibration, Indication and Control Equipment used with Thermocouple Type T | 0.029 °C – 0.0033 % rdg with 0° C RJ 0.13 °C – 0.0083 % rdg with internal RJ |
(-200, -100, 100, 250, 399) °C |
Temperature Simulation use with Thermocouple Type T | 0.015 °C with 0° C RJ 0.13 °C – 0.0067 % rdg with internal RJ |
(-200, -100, 100, 250, 399) °C |
Temperature Calibration, Indication and Control Equipment used with RTD | 0.0020 % rdg + 0.005 °C | (-200, 0, 200, 400, 600, 800) °C |
Temperature Simulation use with RTD | 0.0019 % rdg + 0.005 °C | (-200, 0, 200, 400, 600, 800) °C |
Note: Indications regarding organizational methods, presentation and sending of results will be reported in the ILC participation protocol
Scheduled start date for PT: every year in June
Payment: bank transfer